Title: | Strain-stress analysis of lower limb with applied fixator |
Authors: | Mrázek, Michal Florian, Zdeněk Veselý, Radek Borák, Libor |
Citation: | Applied and Computational Mechanics. 2010, vol. 4, no. 1, p. 67-78. |
Issue Date: | 2010 |
Publisher: | University of West Bohemia |
Document type: | článek article |
URI: | http://www.kme.zcu.cz/acm/index.php/acm/article/view/54/37 http://hdl.handle.net/11025/1391 |
ISSN: | 1802-680X (Print) 2336-1182 (Online) |
Keywords: | deformačně-napěťová analýza;zevní fixace;holenní kost |
Keywords in different language: | strain-stress analysis;external fixation;tibia |
Abstract: | This paper compares physiological state of tibia before and after application of an external fixator. The fixator systems’ models but also model of tibia are loaded in the direction of body axis. The paper is focused on the examination of differences in stiffness before and after the application of fixation. Two types of axial external fixators are compared. Both fixators differ in their construction. The first fixator is two-frame and fixation rods are used for fixing the bone tissue (variant I). The second one is fixed into tibia with screws (variant II). We have found out that the two-frame external fixator has much bigger stiffness during limb fixation than the fixator with one body. Much higher deformations compared to physiological state of tibia occur in the variant II. |
Rights: | © 2010 University of West Bohemia. All rights reserved. |
Appears in Collections: | Volume 4, number 1 (2010) Volume 4, number 1 (2010) |
Please use this identifier to cite or link to this item:
http://hdl.handle.net/11025/1391
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.