Full metadata record
DC FieldValueLanguage
dc.contributor.authorBouřa, Adam
dc.contributor.authorHusák, Miroslav
dc.contributor.editorPihera, Josef
dc.contributor.editorSteiner, František
dc.date.accessioned2012-10-31T10:10:35Z
dc.date.available2012-10-31T10:10:35Z
dc.date.issued2010
dc.identifier.citationElectroscope. 2010, č. 3, EDS 2010.cs
dc.identifier.issn1802-4564
dc.identifier.urihttp://147.228.94.30/images/PDF/Rocnik2010/Cislo3_2010_EDS/r4c3c2.pdf
dc.identifier.urihttp://hdl.handle.net/11025/575
dc.format6 s.
dc.format.mimetypeapplication/pdf
dc.language.isoenen
dc.publisherZápadočeská univerzita v Plzni, Fakulta elektrotechnickács
dc.relation.ispartofseriesElectroscopecs
dc.rightsCopyright © 2007-2010 Electroscope. All Rights Reserved.en
dc.subjectuniverzální zkušební stolicecs
dc.subjectdeformační senzorycs
dc.subjectdálkové měřenícs
dc.titleUniversal test bench for characterization of distance-measuring and strain sensorsen
dc.typečlánekcs
dc.typekonferenční příspěvekcs
dc.typearticleen
dc.typeconferenceObjecten
dc.rights.accessopenAccessen
dc.type.versionpublishedVersionen
dc.description.abstract-translatedFollowing text describes a universal test bench that can be used for characterization of the sensors. It offers sufficient distance and force ability to characterize a wide sort of the sensors. It is suitable for precise distance setting. It is able to set the distance in the range of 10 cm with the resolution up to 10 μm. The mechanical tolerance is negligible. It is less than five micrometers for unloaded viper. Possible relevant tolerance can be caused only by the structure frame bending when the high mechanical stress is applied. The test bench was developed as a research instrument for characterization of new sensor structures and it is also used as a teaching facilitation in the sensor systems tutorial. The test bench functionality is illustrated on five types of the sensors – the capacitive sensor, the optic sensor GP2D120, the Hall sensor A1301, the magneto resistive sensor HMC1501 and for the compression measurement was used the cantilever beam LC501-100.en
dc.subject.translateduniversal test benchen
dc.subject.translatedstrain sensorsen
dc.subject.translateddistance measureen
dc.type.statusPeer-revieweden
Appears in Collections:Číslo 3 - EDS 2010 (2010)
Číslo 3 - EDS 2010 (2010)

Files in This Item:
File Description SizeFormat 
r4c3c2.pdf451,45 kBAdobe PDFView/Open


Please use this identifier to cite or link to this item: http://hdl.handle.net/11025/575

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.