Název: Test bench for calibration of magnetic field sensor prototypes for COMPASS-u tokamak
Autoři: Torres, Andre
Kovařík, Karel
Markovič, Tomáš
Adámek, Jiří
Duraň, Ivan
Ellis, Robert
Jeřáb, Martin
Řeboun, Jan
Turjanica, Pavel
Weinzettl, Vladimír
Fernandes, Horacio
Citace zdrojového dokumentu: TORRES, A., KOVAŘÍK, K., MARKOVIČ, T., ADÁMEK, J., DURAŇ, I., ELLIS, R., JEŘÁB, M., ŘEBOUN, J., TURJANICA, P., WEINZETTL, V., FERNANDES, H. Test bench for calibration of magnetic field sensor prototypes for COMPASS-u tokamak. Fusion engineering and design, 2021, roč. 168, č. July 2021, s. 1-5. ISSN 0920-3796.
Datum vydání: 2021
Nakladatel: Elsevier
Typ dokumentu: článek
article
URI: 2-s2.0-85102246506
http://hdl.handle.net/11025/44920
ISSN: 0920-3796
Klíčová slova v dalším jazyce: magnetic diagnostic;magnetics;Tokamak;mineral insulated cables;MIC;magnetic sensors;frequency response
Abstrakt v dalším jazyce: Magnetic field sensors are fundamental for control and physics exploitation of fusion devices. Their inductive nature implies high dynamic ranges and a broad bandwidth, thus a precise characterization and calibration of these probes is paramount. COMPASS-U will have a completely new set of magnetic diagnostics, from sensors to data acquisition. Sensors installed in-vessel will operate at 300–500 ◦C and should survive transients of even higher temperatures and thus materials used should be high-temperature compatible. These design limitations will have an impact on the dynamic range and bandwidth of the sensors, which needs to be reliably quantified, optimally with the same test bench for all sensor types. The first part of this work presents a test bench and the process of how to calibrate the effective area using a large solenoidal coil and frequency response of the magnetic sensor prototypes with a Helmholtz coil. In the second part, test results of the sensor prototypes are presented and discussed in detail. The low-bandwidth sensors made of mineral insulated cable (MIC), intended for plasma control and machine protection, show negligible attenuation up to 10 kHz, sufficient for their role. For fast coils consisting of bare wire wound on ceramic mandrel and Thick Printed Copper (TPC) sensors the negligible attenuation measured below 1 MHz is again sufficient for their intended purpose of detecting fast coherent plasma fluctuations. Resonances introduced by the capacitance of long cables from the vacuum vessel feedthroughs to the data acquisition systems are measured, to model their influence on the signal.
Práva: Plný text je přístupný v rámci univerzity přihlášeným uživatelům.
© Elsevier
Vyskytuje se v kolekcích:Články / Articles (KET)
Články / Articles (RICE)
OBD

Soubory připojené k záznamu:
Soubor VelikostFormát 
Reboun_1-s2.0-S092037962100243X-main-1.pdf2,36 MBAdobe PDFZobrazit/otevřít  Vyžádat kopii


Použijte tento identifikátor k citaci nebo jako odkaz na tento záznam: http://hdl.handle.net/11025/44920

Všechny záznamy v DSpace jsou chráněny autorskými právy, všechna práva vyhrazena.

hledání
navigace
  1. DSpace at University of West Bohemia
  2. Publikační činnost / Publications
  3. OBD